Indian J Anaesth. 2018 Sep; 62(9): 691–697.

Approach to failed spinal anaesthesia for caesarean section

Ketan S Parikh

oneDepartment of Anesthesia, Alienation Candy Infirmary, Mumbai, Maharashtra, Republic of india

2Department of Anesthesia, Mumbai Hospital and Medical Research Eye, Bombay, Maharashtra, India

Shwetha Seetharamaiah

3Section of Anesthesia, Janani Anesthesia and Critical Care Services, Shimoga, Karnataka, India

Abstract

Failure of spinal amazement for caesarean section may have deleterious consequences for the mother too as the newborn baby. In this article, we discuss the mechanisms of failure of spinal amazement likewise every bit the arroyo to a failed cake. Nosotros performed a literature search in Google Scholar, PubMed, and Cochrane databases for original and review manufactures apropos failed spinal anaesthesia and caesarean section. Strategies for a failed spinal anaesthetic include manoeuvers to salvage the block, repeating the block, epidural anaesthesia or a combined spinal–epidural (CSE) technique, or resorting to general amazement. Factors influencing the selection of these alternative options are discussed. A "failed spinal algorithm" tin guide the anaesthesiologist and help reduce morbidity and mortality.

Key words: Caesarean section, failed spinal, general anaesthesia, repeat spinal, safety practise

INTRODUCTION

Neuraxial anaesthesia is the commonest, safest, and virtually logical choice of anaesthesia for caesarean section. Diverse options include spinal amazement, epidural anaesthesia, or combined spinal epidural (CSE) anaesthesia. Of all these, single-shot spinal is the nearly widely skillful fashion of amazement in both elective equally well every bit emergency situations. Although spinal anaesthesia is considered to be the most reliable form of anaesthesia, occasional failures are not unknown. Information technology is essential to recognize that the failure of spinal anaesthesia during caesarean section has detrimental implications on the wellbeing of both the parturient and the neonate. Therefore, the failure must be viewed more than critically in obstetric compared with nonobstetric settings. We performed a literature search in Google Scholar, PubMed, and Cochrane databases for original and review manufactures concerning failed spinal anaesthesia and caesarean department.

The Saving Mothers Report,[1] which assessed the deaths of 92 parturients in Southward Africa, between 2008 and 2010, revealed that 73 (79%) patients died due to spinal anaesthesia. Out of these, ten deaths were related to the complications of a subsequent general anaesthesia administered when spinal anaesthesia proved inadequate for surgery. Lack of clinical experience and inappropriate approach to failure were responsible for maternal mortality. Every bit there are very limited options to approach the failure, utmost vigilance is warranted while performing spinal anaesthesia to minimise both failure rate also every bit maternal or foetal complications. With careful functioning of technique, a failure rate as depression as 1% is accessible though various studies have quoted failure rate upwardly to 17%.[2,iii]

WHAT CONSTITUTES FAILURE

Broadly speaking, pain or discomfort during caesarean section necessitating the need to undertake additional measures to go on surgery is regarded as failed spinal anaesthesia. The failure can be defined equally "failure to provide satisfactory surgical weather condition and/or maternal comfort and satisfaction during caesarean section with or without conversion to general anaesthesia."

MECHANISMS OF FAILURE

As early as in 1922, Gaston Labat made a statement:[4] "Two conditions are, admittedly necessary to produce spinal anaesthesia; Puncture of the dura mater and Subarachnoid injection of an anaesthetic agent." Inability to achieve these two master goals due to any cause leads to failure of spinal anaesthesia. The failure could be operator, technical, or equipment related [Table 1]. Meticulous attention to the following ten points will help reduce the failure rate of spinal anaesthesia.

Table 1

Mechanism of failure

An external file that holds a picture, illustration, etc.  Object name is IJA-62-691-g001.jpg

Dural puncture

Disability to either puncture the dura (dry tap) or obtain gratuitous catamenia of cerebro-spinal fluid (CSF) later declared dural puncture is ane of the obvious causes of failure of spinal anaesthesia. The primary reasons are blocked needle, poor patient positioning, and faulty needle placement technique. Obesity, anatomical abnormalities, and an broken-hearted patient will frequently add together to the misery.

Information technology is prudent to bank check the patency of the needle prior to insertion and not to advance the needle without the stylet in identify. Bent or crooked needles should be discarded. The patient should be positioned on a firm flat surface with maximal flexion in order to open the lumbar spine, fugitive lateral curvature of the spine. Though the sitting position is often easier, there is no departure betwixt the lateral and sitting positions in terms of failure charge per unit.[5] The take a chance of traumatic or failed spinal block may exist reduced past the apply of ultrasound guidance.[6]

Pseudosuccess of dural puncture must be identified. This is most likely due to excess local anaesthetic solution infiltrated before the procedure or attempting spinal after a prior epidural analgesia has failed. Very rarely, puncture of a congenital arachnoid cyst may yield CSF-similar fluid.[7]

Intrathecal degradation of the entire volume

Leakage of injected volume tin occur due to poor connexion between the needle and the syringe. Ensuring that the needle tip does not get displaced while trying to firmly secure the syringe or during drug injection tin can forestall the injectate loss during drug administration. Routine aspiration of CSF earlier injecting and only before completion of injection of local anaesthetics is one way of confirming intrathecal delivery of the drug. Use of finer pencil point needles can exist technically challenging as CSF aspiration through effectively needles, for example, 29 G is difficult. If this needle with side opening is placed incorrectly across the dura, part of the injectate will be lost in the epidural or subdural space. Sometimes, a small dural flap formed during the procedure may act equally a "flap" valve. CSF can exist aspirated freely merely during injection, an inward displacement of the flap leads to loss of injectate between the dura and the arachnoid mater.[8]

Injection of the correct drug

Prevention of injection of wrong drug is of utmost importance. Apart from failure, wrong drug injection tin can crusade maternal mortality (e.grand., tranexamic acid.).[ix] Such errors are unpardonable and must exist avoided by double checking of drugs earlier injection, use of prefilled syringes, and use of luer-lock connection.[10]

Injection of the correct dose and volume

The speed of onset, quality, and elapsing of the block are solely determined by the dose of the local anaesthetic. The spread and therefore the level of sensory block is adamant by the density/baricity of the solution, the local anaesthetic volume, and patient's position.[eleven] The spread of isobaric local anaesthetic is far more than unpredictable than hyperbaric solutions used routinely. Hypobaric solution has potential do good when injected in the sitting position, simply not without the risk of hypotension. Use of adjuvants may hasten the onset, prolong the elapsing, and improve the quality of motor cake merely may not prevent failure. According to the study by Ginosar et al.,[12] ED50 of bupivacaine is 7.five mg and ED95 (surgical success) is 11 mg. Anatomical factors like Kyphosis and/or scoliosis are responsible for the technical difficulty and interference with spread. Unilateral block may be caused by supporting ligaments within the theca acting as a longitudinal barrier. Very rarely, inadequate spread or failure may exist explained by asymptomatic neurological pathology within the vertebral canal.[13]

Facilitating action of the drug on the nerve roots

The efficacy and potency of the drug can get altered by various factors such as prolonged exposure to sunlight, excessive dilution of the drug, chemical incompatibility afterward mixing with other drugs, or altered pKa due to interaction with the alkaline metal CSF. In such conditions, fifty-fifty if the unabridged drug reaches the nerves, the desired activity may not be obtained. Mixing of two drugs may crusade precipitation or lowering of the pH altering the movement of local anaesthetics across the neuronal membrane. Local anaesthetic resistance due to mutation of sodium channel has been reported as a cause of ineffective drug action in few patients.[14]

Enough time for the cake to found

An extremely anxious parturient who perceives every stimulus every bit pain and an anaesthesiologist who does not let adequate fourth dimension for the block to establish tin precipitate pseudofailure of spinal amazement. Sufficient time should exist given for the cake to establish before labeling information technology as failed spinal. However, the maximum fourth dimension limit for the onset of action later on drug deposition into the intrathecal space is 15–xx minutes. The drug is unlikely to produce desired action after this fourth dimension menstruation.[eight] The slower the onset of either sensory or motor component of the cake, the more likely an inadequate block will result.

Assessment of the cake before assuasive the skin incision

Formal testing of the block prior to showtime of surgery is the key to success. The level of the block must be checked bilaterally and documented properly on the amazement chart without neglect. At that place is no consensus as to the best exercise nearly checking the block. However, three modalities like, sensation of cold (ice cubes or ethyl chloride spray), light touch (cotton fiber swab), and loss of motor power are used commonly.[fifteen] Use of pinprick method is non recommended. It must be understood that adequate level of the cake does NOT guarantee its quality. Although level of block needed for abolishing somatic pain during caesarean section is T10 dermatome, a block as high as T4 is required to abolish visceral pain and discomfort.[sixteen,17] Cold sensation felt at T4 or lower, calorie-free touch sensation felt at T8 or lower, and poor motor block in lower extremities later on 10 minutes are causes of concern virtually capability of the block.

Postprocedure position

A slight head low position with 15° left lateral tilt allows acceptable spread of the local anaesthetic avoiding aortocaval compression. Higher spread may exist accomplished by turning patient on left side with head down, to obliterate lumbar lordosis. Maneuvers similar flexing the hip or knee–chest position may non be feasible.

Communication skills

Communication before and during the procedure with the parturient explaining the need for correct positioning and cooperation is very vital. If the patient is in agile labor, it is very difficult for her to cooperate and maintain proper position. I must allay their anxiety of pain during the procedure by informing about local infiltration before the actual injection. Autonomously from proper positioning of the parturient after the block and testing of the level of anaesthesia before the skin incision, positive communication with the broken-hearted parturient is vital in order to prevent pseudofailure. The onset of level of anaesthesia should not be assessed likewise soon or too frequently every bit information technology might innovate dubiety or anxiety in the mind of the parturient. Patient'southward expectations play a major role and if their anxiety and concerns are not allayed, they are more than probable to merits that the anaesthesia has non worked leading to increased litigation.[8]

Personal feel of the clinician

Spinal anaesthesia with 25/27 or 29 Yard spinal needles take relatively higher failure rate due to technical issues that tin can be improved upon only with practice. An experienced anaesthesiologist volition practise proper preoperative assessment, bank check the equipment, and perform the procedure in a at-home, unhurried mode, winning the patient's confidence and cooperation, and hence improving success rate.[xiv]

TYPE OF FAILURE

Clinically, the failure may range from no block at all to failure despite a proper block. There can be complete absence of both sensory and/or motor block or a good sensory motor cake with inadequate cephalad spread, unsuitable for surgical anaesthesia. One may likewise come across unilateral cake either because of positioning or anatomical bulwark due to presence of longitudinal ligament blocking the fifty-fifty spread. Inadequate or partially misplaced dose may result in patchy block or reduced duration of cake causing failure of the block to terminal the duration of surgery.

HOW TO APPROACH INADEQUATE REGIONAL BLOCK

Failure of a spinal anaesthetic is associated with serious clinical, psychological, and medico-legal consequences, especially if the failure becomes evident after starting the surgery. The Majestic College of Anaesthetists, UK, propose that the conversion rate from spinal to general anaesthesia should exist <1% in elective and <iii% in emergency caesarean section.[18] While dealing with inadequate block, 1 must remain calm and appear to be in total command. Frequent testing may increase feet and result in loss of rapport. Save of the block should be the priority avoiding the general anaesthesia as far every bit possible. If parturient complains of hurting after the surgery has started, surgery should exist stopped immediately. Further steps involve reassuring the patient, assessing the type of failure, and trying to maximize the comfort with the all-time alternative available under the circumstances. Table 2 summarizes the various strategies to deal with a failed spinal during caesarean section.

Table 2

Management options; 3Rs

An external file that holds a picture, illustration, etc.  Object name is IJA-62-691-g002.jpg

OPTIONS FOR Direction OF FAILURE

Revive or salvage the cake

Prior to initiation of surgical procedure, the failing block can be revived by concrete manoeuvers similar placing the patient on her side (left lateral) along with caput low, limited hip flexion to straighten the back and obliterate the spine curvature, and valsalva manoeuver, or coughing (epidural volume expansion), to facilitate cephalad spread. If an epidural catheter is in identify, cephalad spread may exist achieved with additional doses of local anaesthetics or saline [Tabular array 3].

Table 3

Measures to revive the block

An external file that holds a picture, illustration, etc.  Object name is IJA-62-691-g003.jpg

After the surgery has started, the options are limited. Systemic sedation and/or analgesia with intravenous opioid (fentanyl), or anxiolysis (midazolam) may be helpful for the breakthrough pain. Inhalation of entonox or local infiltration of the wound may be an option to consider depending upon the situation. Sedation leading to loss of consciousness, or general anaesthesia without securing the airway is Non RECOMMENDED. If more than than one option to salve the block is unsuccessful, have failure.

Repeat the regional block

For a failed spinal amazement, repeating the block is a sensible option if feasible. The ease with which the dura was punctured in the kickoff endeavor should guide the clinician to decide whether to consider the patient for repeat spinal.[19] Afterward waiting for xx minutes to avoid pseudofailure, a dissimilar interspace must exist tried to avert the anatomic distortions. If no effect is seen subsequently twenty minutes post-obit the injection, information technology seems reasonable to repeat the block with full dose. In fractional failure, there must exist some amount of local anaesthetic in the CSF; hence, reduction in the dose by 25%–30% may be necessary if the block is close to T10 [Table 4].[8,20] If the circumstances let and expertise is available, consider using a CSE technique with a reduced intrathecal dose and epidural catheter as backup.

Table four

Repeating the cake

An external file that holds a picture, illustration, etc.  Object name is IJA-62-691-g004.jpg

Evidence in the literature regarding safety of repeat spinal anaesthetic is conflicting.[21,22] Repeating spinal anaesthesia later a failed neuraxial block is not without concerns. Unpredictable reliability and quality of the resulting block may atomic number 82 to a high or total spinal anaesthesia. Sometimes, the repeat cake worsens the haemodynamics without improving the block. Exposure of high concentration of drugs on the nerve roots due to restricted spread may cause cauda equina syndrome. Multiple attempts can increase gamble of postdural puncture headache or vascular injury leading to epidural haematoma. As the side by side nerve tissue is already affected by local anaesthetic action, risk of nerve damage due to straight needle trauma is a theoretical possibility [Tabular array v]. If the echo spinal fails to produce desired block, general amazement is the safest option.

Table 5

Complication of repeat spinal injection

An external file that holds a picture, illustration, etc.  Object name is IJA-62-691-g005.jpg

Failed spinal in combined spinal epidural

Inadequate spinal block in CSE can exist approached by injecting saline or incremental doses of local anaesthetic through the epidural catheter (volume expansion theory) to increase the block meridian past squeezing the intrathecal space.[23] If spinal cake is not credible within 15–20 minutes of injection, epidural catheter tin be used to establish surgical anaesthesia.

Recourse to full general anaesthesia

The recourse to full general anaesthesia should be done after a prompt and through assessment of the situation using common sense and clinical experience, without compromising parturient' safety or comfort. Converting to full general anaesthesia in a parturient is not without risks like hypotension on consecration, aspiration, and potentially difficult airway, particularly under unfavorable environment when dealing with foetal distress, or if failure becomes apparent after the pare incision.

CHOOSING APPROPRIATE OPTIONS FOR FAILED BLOCK

Primary factors to consider while approaching failure are as follows: first, if the skin incision is already fabricated, and second, how urgent is information technology to deliver the infant. Communication and cooperation between the anaesthesiologist and obstetrician is crucial in choosing safe anaesthesia technique for the parturient.[24] Starvation status, difficult airway, ease of performance of regional block, and comorbid condition of the parturient are secondary factors to be considered while choosing the right pick to arroyo the failure.[19] A uncomplicated algorithm as described in Effigy i may be followed to choose the best possible option under the circumstances.

An external file that holds a picture, illustration, etc.  Object name is IJA-62-691-g006.jpg

Algorithm: safety approach to failed spinal anaesthesia.

#Assess: Hard Airway, Starvation Condition, Technical Difficulty in performing SA, Co-morbidity. GA=General amazement

In case of failure earlier skin incision is made and in that location is no urgency to deliver in <30 minutes, repeating the regional cake may be the safest option for the parturient

In case of failure before peel incision is made and there is urgency to deliver in <30 minutes, make a quick assessment of state of affairs. If performing the spinal cake has been easy and the patient is cooperative, quick attempt at repeating a spinal may be considered. If that is not viable, try to salve the cake while getting gear up for recourse to general anaesthesia without any further delay. Avoid whatsoever compromise on maternal or foetal safety.

In case of failure subsequently pare incision is fabricated and there is no urgency to deliver in <30 minutes, revive the block with various options appropriate to the circumstances. Though attempts should be made to avert general anaesthesia, especially in anticipated difficult airway, information technology may be the just selection to safely carry on surgery without further filibuster once the skin is already incised. In this situation, a positive decision on the choice of anaesthesia must always be made prior to the uterine incision.

In case of failure later skin incision is made and there is urgency to deliver in <xxx minutes, recourse to general Anaesthesia in the best interest of maternal and foetal safety.

Fundamental POINTS TO Rubber PRACTICE

ABC of safe practice is summerised in Table 6 below;

Table 6

Central points to safe practice

An external file that holds a picture, illustration, etc.  Object name is IJA-62-691-g007.jpg

  1. Avoid excessive and multimodal sedation: "A Bit of this plus a Bit of that is a classic recipe for DISASTER!"

  2. Be prepared: Be honest to yourself and your patient, have failure, be ready to convert to general anaesthesia

  3. Communication: Reassure and explain the state of affairs. If possible, include the parturient in the discussion nigh how to continue in case of failure giving various options

  4. Documentation: Meticulous documentation of the testing and level of the cake achieved. Every effort at reviving the block, including reasons why particular decisions were made, and its effectiveness must be documented. When general anaesthesia is offered, the timing along with patient'south response must be recorded clearly in notes, especially if it was declined past patient

  5. Explanation and postoperative follow-up: Provide an honest caption of what happened and reply any questions or concerns they may accept, including implications for future pregnancy. Proper documentation and good communication are keys to foreclose medico-legal problems.

SUMMARY

The recipe for successful spinal anaesthesia is "correct dose, correct drug, correct place." Failed spinal anaesthesia is associated with significant maternal morbidity and mortality. Due to restricted rescue options to salvage the cake, meticulous attention to forestall the failure is prudent. The first choice in dealing with the failure is to salvage the block without hastily converting it to general anaesthesia. Choice of inappropriate selection or overzealous utilise of systemic analgesia and sedation to approach failure tin can dangerously compromise parturient safety. Despite controversies, repeat intrathecal injection performed by an experienced anaesthesiologist nether vigilance is a very good option. A "failed spinal" algorithm guides the anaesthesiologist to approach "failures," reduces maternal and foetal morbidity and bloodshed, and addresses pursuant litigation.

Fiscal back up and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

REFERENCES

i. Africa GoS. Saving Mothers. Pretoria: Department of Health; 2012. National Commission on Confidential Enquiries into Maternal Deaths. [Google Scholar]

2. Levy JH, Islas JA, Ghia JN, Turnbull C. A retrospective study of the incidence and causes of failed spinal anesthetics in a university infirmary. Anesth Analg. 1985;64:705–ten. [PubMed] [Google Scholar]

3. Sng BL, Lim Y, Sia AT. An observational prospective cohort study of incidence and characteristics of failed spinal anaesthesia for caesarean section. Int J Obstet Anesth. 2009;xviii:237–41. [PubMed] [Google Scholar]

iv. Labat One thousand. Regional Anesthesia: Its technique and Clinical Applications. Philadelphia: WB Saunders Company; 1922. [Google Scholar]

5. Inglis A, Daniel M, McGrady E. Maternal position during induction of spinal anaesthesia for caesarean section. A comparison of right lateral and sitting positions. Amazement. 1995;50:363–5. [PubMed] [Google Scholar]

six. Ghosh SM, Madjdpour C, Mentum KJ. Utrasound-guided lumbar central neuraxial block. BJA Educ. 2016;16:213–20. [Google Scholar]

7. Stace JD, Gaylard DG. Failed spinal anaesthesia. Anaesthesia. 1996;51:892–3. [PubMed] [Google Scholar]

8. Fettes PD, Jansson JR, Wildsmith JA. Failed spinal anaesthesia: Mechanisms, management, and prevention. Br J Anaesth. 2009;102:739–48. [PubMed] [Google Scholar]

9. Garcha PS, Mohan CV, Sharma RM. Death after an inadvertent intrathecal injection of tranexamic acrid. Anesth Analg. 2007;104:241–two. [PubMed] [Google Scholar]

10. Kinsella SM. The daze of the 'nuer': The Britain experience with new non-luer neuraxial equipment and implications for obstetric anaesthesia. Int J Obstet Anesth. 2013;22:1–5. [PubMed] [Google Scholar]

11. Carpenter RL, Hogan QH, Liu SS, Crane B, Moore J. Lumbosacral cerebrospinal fluid volume is the principal determinant of sensory block extent and duration during spinal anesthesia. Anesthesiology. 1998;89:24–9. [PubMed] [Google Scholar]

12. Ginosar Y, Mirikatani E, Drover DR, Cohen SE, Riley ET. ED50 and ED95 of intrathecal hyperbaric bupivacaine coadministered with opioids for cesarean delivery. Anesthesiology. 2004;100:676–82. [PubMed] [Google Scholar]

xiii. Kavlock R, Ting PH. Local anesthetic resistance in a pregnant patient with lumbosacral plexopathy. BMC Anesthesiol. 2004;4:1. [PMC free article] [PubMed] [Google Scholar]

14. Drasner Yard. Spinal anaesthesia: A century of refinement, and failure is still an option. Br J Anaesth. 2009;102:729–30. [PubMed] [Google Scholar]

15. Russell IF. A comparing of cold, pinprick and bear upon for assessing the level of spinal cake at caesarean section. Int J Obstet Anesth. 2004;13:146–52. [PubMed] [Google Scholar]

16. Russell IF. Assessing the block for caesarean section. Int J Obstet Anesth. 2001;10:83–5. [PubMed] [Google Scholar]

17. Riley ET. Regional anesthesia for cesarean department. Tech Reg Anesth Pain Manage. 2003;7:204–12. [Google Scholar]

xviii. Kinsella M. Raising the Standard: A Compendium of Audit Recipes. London: The Royal College of Anesthetists; 2012. [Google Scholar]

19. Parikh KS, Pandya ST. World Clinics: Anesthesia, Critical Care and Pain – Analgesia and Anesthesia in Labor and Delivery. Vol. 1. New Delhi: Jaypee Medical Publishers Ltd.; 2013. Management of failed spinal anesthesia for lower segment cesarean section; pp. 457–71. [Google Scholar]

twenty. Deshpande Due south, Idriz R. Repeat dose after an inadequate spinal block. Anaesthesia. 1996;51:892. [PubMed] [Google Scholar]

21. Pokharel A. Written report of failed spinal anesthesia undergoing caesarean section and its management. Postgrad Med J NAMS. 2011;11:11–5. [Google Scholar]

22. Drasner K, Rigler ML. Echo injection after a "failed spinal": At times, a potentially unsafe do. Anesthesiology. 1991;75:713–4. [PubMed] [Google Scholar]

23. Beale N, Evans B, Plaat F, Columb MO, Lyons Yard, Stocks GM, et al. Outcome of epidural volume extension on dose requirement of intrathecal hyperbaric bupivacaine at caesarean section. Br J Anaesth. 2005;95:500–iii. [PubMed] [Google Scholar]

24. Dyer RA, Reed AR, James MF. Obstetric anaesthesia in depression-resource settings. All-time Pract Res Clin Obstet Gynaecol. 2010;24:401–12. [PubMed] [Google Scholar]